MSgreen
SQUID_MSgreen2.jpg
magnetometer chips
The model MSgreen is designed for extremely sensitive measurements of small magnetic fields such as biomagnetic fields of the human heart or brain.
With a field sensitivity of better than 3.5 fT/Hz1/2 MSgreen is well-suited for these types of applications.

The sensor is housed and sealed inside a cylindrical holder (diameter 1mm, mm high) made of glass fibre reinforced plastic. The sensor chip integrates an heater for de-fluxing.

Summary of important parameters:  
 
Characteristics 
Chip size in mm 7.5×7.5
Magnetometer effective area Aeff = 2.4 mm²; 1/Aeff =0.86nT/Ф0
Voltage modulation more than, in µV 40
White noise level better than, in µФ0/Hz1/2 4
Working temperature in K 1...5

 
Three feedback coils available with a mutual inductance of:  
 
M= 70 pH 30µA/Ф0
M= 40 pH 50µA/Ф0
M= 35 pH 60µA/Ф0

 

« back

Contact us

Supracon AG
An der Lehmgrube 11
07751 Jena
Germany

Tel.: +49-3641-2328100
Fax.: +49-3641-2328109

info(at)supracon.com