Typical calibration accuracy (direct comparison to a second Josephson Voltage Standard)
± 4 nV @ 10 V
ΔV/V10 V = 4×10-10
Typical calibration accuracy of secondary voltage standards (limited by the noise of the secondary voltage standard)
± 20 nV @ 1V
ΔV/V1 V = 2×10-8
± 100 nV @ 10 V
ΔV/V10 V = 1×10-8
Thermal voltage of wires and reversal switch
< 10 nV @ all 3 channels
Typical gain factor g of external voltmeter (depends on the type of voltmeter)
Δg/g < 3×10-7
Direct comparison of supraVOLTcontrol with the Josephson Voltage Standard of the Physikalisch-Technische Bundesanstalt (PTB) @ 10 V.
The measured voltage difference corresponds to an accuracy of 7×10-11 with a measurement uncertainty of 3.4×10-10.