AC voltage standard array
chip_10v.jpg
The centrepiece of the AC Quantum Voltmeter is a 10 Volt programmable Josephson voltage standard circuit

  • Number of Josephson junctions: 69632
  • Maximum output voltage: ±10.08 V
  • Operating frequency: 70 GHz
  • Zero & first order Shapiro step: 1 mA
  • Bias current: ±6 mA
  • Voltage increment: 145 µV  
 
V = n × f / KJ90
V Josephson voltage
KJ90
Josephson constant
n programmable integer
f
microwave frequency

 

« back